| Model Number |
TRM-310 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/21/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported left side exchange from textured to smooth.Healthcare professional later reported capsular contracture, baker grade unknown.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture, baker grade unknown" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: lab analysis of the returned device identified: weight to the spec, deformation, white particles on the surface of the shell.Cloudy was observed after autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.Additional, changed, and/or corrected data.
|
| |
|
Event or Problem Description
|
|
Device has been replaced.
|
| |
|
Search Alerts/Recalls
|