• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in a fistula of a vessel.A 7.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: a gladiator device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit and positive pressure was applied and liquid was observed to be leaking from a balloon pinhole located approximately 3mm distal of the proximal marker band.A visual examination observed no damage to the tip of the device.A visual and tactile examination of the shaft of the device found no issues.A visual examination found no issues or damage to the markerbands of the device.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in a fistula of a vessel.A 7.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.It was further reported that the balloon was inflated once at less working pressure for a few seconds before it ruptured.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in a fistula of a vessel.A 7.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.It was further reported that the balloon was inflated once at less working pressure for a few seconds before it ruptured.
 
Additional Manufacturer Narrative
Updated b5: describe event or problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13866084
Report Number2134265-2022-03309
Device Sequence Number4716813
Product Code LIT
UDI-Device Identifier08714729809791
UDI-Public08714729809791
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/26/2022
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0023856775
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer03/22/2022
04/15/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date04/06/2022
04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
Patient Weight60 KG
-
-