| Model Number |
1897200 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis found motor shorted and bearings corroded.Tool cannot be inserted and can't perform pre-temperature test.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care professional (hcp) reported via distributor that the handpiece overheated during operation.There is no patient impact.Upon follow up, it was reported that the device overheated in less than a minute.They turned on the power, connected the handpiece to the console, stepped on the pedal, and it got hot during use.There was no delay in the surgical procedure but a back up device was used.
|
| |
|