Catalog Number 8065751510
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Uveitis (2122); Toxic Anterior Segment Syndrome (TASS) (4469); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported that after a cataract surgery patient experienced inflammation (fibrin) and endophthalmitis.Patient was given with an antibacterial agent and the condition was improving.It was informed that the patient was in contact with surgical procedure packs, viscoelastics, ophthalmic phacoemulsification tip, ophthalmic phacoemulsification handpiece and irrigation/aspiration tip.Additional information was received clarifying the surgeon confirmed the event was tass (toxic anterior segment syndrome).
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of inflammation post op, toxic anterior syndrome (tass), and medical intervention; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A review of the sterilization records indicates that the product was sterilized and released according to the product's acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All irrigation/aspiration (i/a) tips are 100% visually inspected prior to being released from the manufacturing facility.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS
Common Device Name
DEVICE, IRRIGATION, OCULAR SURGERY
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13867697
Report Number2523835-2022-00106
Device Sequence Number4715833
Product Code KYG
UDI-Device Identifier00380657515103
UDI-Public00380657515103
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number8065751510
Device Lot Number13XW21
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION SURGICAL PROCEDURE PAK; CONSTELLATION SURGICAL PROCEDURE PAK; CONSTELLATION VISION SYSTEM; CONSTELLATION,TABLETOP; DISCOVISC OPHTHALMIC VISCOSURGICAL DEVICE; HOYA XY1AT4; INFINITI VISION SYSTEM, HANDPIECE TIP; MONARCH III IOL DELIVERY SYST; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; OZIL TORSIONAL HANDPIECE
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age81 YR
Patient SexFemale
-
-