• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE Back to Search Results
Catalog Number 03P76-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On 15-mar-2022, abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded a suspected discrepant potassium result of 5.0 mmol/l on a 72 year old female patient in a coma, and end-stage chronic renal failure.There was no additional patient information available at the time of this report.Return product is not available for investigation.Method, date, collected, tested, results, sample; i-stat, (b)(6) 2022, 10:10, 10:14, 5.0 mmol/l, a; lab, (b)(6) 2022, 09:28, 10:28, 6.8 mmol/l, b.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 06-apr-2022.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained testing met the acceptance criteria found in q04.01.003 rev.Ag, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for eg7+ cartridge lot n22028.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT EG7+ CARTRIDGE
Common Device Name
EG7+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key13869189
Report Number2245578-2022-00058
Device Sequence Number5295616
Product Code CHL
UDI-Device Identifier10054749000040
UDI-Public10054749000040
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K940918
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2022
Device Catalogue Number03P76-25
Device Lot NumberN22028
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date03/23/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-