| Model Number |
CAX400T12 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Headache (1880); Cough (4457)
|
| Date of Event |
03/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam.The patient alleged visualization of a yellow cloud in the hose.Patient also alleged a headache and cough.Patient also alleged the device is noisy, running hot and burned the table.There is no allegation of serious or permanent harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices. the manufacturer previously received information alleging visualization of a yellow cloud in the hose and also alleged a headache and cough.Patient also alleged the device is noisy, running hot and burned the table related to a cpap device's sound abatement foam.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting an updated report at this time.If pertinent information becomes available to the manufacturer at a later date, a follow-up report will be filed.Section h6 updated in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging visualization of a yellow cloud in the hose and also alleged a headache and cough.Patient also alleged the device is noisy, running hot and burned the table related to a cpap device's sound abatement foam.The dreamstation cpap pro device was returned to a service center for further evaluation.During the evaluation of the dreamstation cpap pro device at the service center, sound abatement foam degradation was not observed.The technician indicated that there was no evidence of foam degradation or particles in the assembly.The technician observed dust contamination.A heat test was performed on the device to verify that the temperatures were within the required perimeters.The test results showed the device showed temperatures within the required threshold.The service center retrieved and reviewed the device's error logs.There was 0 errors found.The service center could not confirm the customer's allegation and there was no finding of visible foam degradation.
|
| |
|
Search Alerts/Recalls
|