| Catalog Number |
07028148190 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump was damaged ship pump and return pump.No harm requiring medical intervention was reported.The insulin pump was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Clarification was received that the questionable results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration signals were higher than expected.Qc was acceptable.The recommended instrument maintenance was completed.Sample alarms were observed in the customer's alarm trace data.The investigation did not identify a product problem.The cause of the event could not be determined.Based on the calibration and qc data, a general reagent issue can be excluded.Product labeling states: "for diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.".
|
| |
|