| Model Number |
7345 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Awaiting product return to smiths medical.If the product is returned, smith medical will return this complaint for further investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during usage the device leaked.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6: updated.H10: device evaluation: the device was returned for investigation.Visual inspection and functional test were performed.The device was found to have a crack in the filter.Upon function testing the reported problem could not be duplicated.The root cause of the reported problem was unknown as the device passed function testing.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture., corrected data: g5.
|
| |
|