(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00073 and 0002648920-2022-00060.Medical devices: articular surface medial congruent (mc) right 14 mm height catalog#: 42522100814 lot#: 64645156.All-poly patella cemented 29 mm diameter catalog#: 42540200029 lot#: 64775971.Tibia cemented 5 degree stemmed right size e catalog#: 42532007102 lot#: 64773559.Stem extension tapered cemented 14 mm diameter +30 mm length catalog#: 42557000114 lot#: 64722097.Foreign source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, and h10.Reported event was confirmed by review of x-rays.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: lateral patellar dislocation is seen, no evidence of implant wear, fracture, or loosening, tibial and femoral implants are anatomically aligned, bone quality is osteopenic.Device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issue for this item and the reported part and lot combination.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|