MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE; PISTON SYRINGE Back to Search Results
Model Number 324910
Device Problem Inaccurate Information (4051)
Patient Problem Hypoglycemia (1912)
Event Date 03/08/2022
Event Type  malfunction  
Event Description
It was reported when using the bd insulin syringe with the bd ultra-fine¿ needle, the device experienced scale marking issues.This event occurred 160 times.The following information was provided by the initial reporter.The customer stated: the very first scale mark (zero point) is marked the way too lower than it should be.Therefore, it makes to draw more insulin.After injection, glucose level was drop to 50mg/dl (more than usual).
 
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: samples were received and an investigation was performed.Bd was not able to duplicate or confirm the indicated issue hence the root cause is undetermined.Unable to perform complaint lot history check for glucose level due to unknown lot number.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13871415
MDR Text Key287826571
Report Number1920898-2022-00187
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00382903249107
UDI-Public00382903249107
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number324910
Device Catalogue Number324910
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-