• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. EB EXC GRN 36IN 2-0 D/A V-7; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. EB EXC GRN 36IN 2-0 D/A V-7; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number X966H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a hernia repair procedure on (b)(6) 2022 and suture was used.During the procedure, while attempting to tie a knot during the fundoplication portion of the procedure the suture broke.Another like device was used to complete the procedure.No adverse patient consequences were reported.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Device analysis: one packet of product code x966h was returned for analysis.Upon initial inspection of the sample, no external damages were observed on the packet.In order to evaluate the conditions of the returned sample, the packet was opened, and no defects were detected.The swage and attachment area was noted to be as expected.The suture wad dispensed without problems and examined along the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.Functional test was performed, and the tensile strength result was above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB EXC GRN 36IN 2-0 D/A V-7
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13872845
Report Number2210968-2022-02116
Device Sequence Number4520072
Product Code GAT
UDI-Device Identifier10705031058477
UDI-Public10705031058477
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberX966H
Device Catalogue NumberX966H
Device Lot NumberRJMCET
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-