| Model Number |
115-222 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "a right side rupture as well as bilateral exchange from textured to smooth due to concern with the product." device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "a right side rupture as well as bilateral exchange from textured to smooth due to concern with the product." device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual analysis of the returned device identified one extended opening and fold creases.A weight test of the device was verified and the device underweight.A microscopic analysis was performed which identified one opening sharp.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: one sharp opening in the side radius assessed as unidentified (tear) opening.
|
| |
|