Model Number 83779
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2022.It was reported that the coil could not be advanced into the microcatheter.A 22mm x 60cm interlock was selected for use.During the procedure, it was noted that the coil could not be advanced into the microcatheter.The procedure was completed with another of the same device.There were no patient complications reported and patient was stable.However, device analysis revealed that there was a missing fiber bundles.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The main coil, pusher wire and introducer sheath were returned for this complaint.The main coil and the pusher wire were interlocked within the introducer sheath.The introducer sheath was inspected, and no damages were noted.The twist lock was opened and the pusher wire was found to be kinked.The main coil was also found kinked and stretched.The delivery/pusher wire was advanced through the introducer sheath; however, it was not possible to continue advancing, as the coil was stuck.The introducer sheath was cut, in order to release the main coil.The zap tip was inspected and had a smooth surface.The interlocking arm was inspected, and no anomalies was noticed.The proximal end has a smooth surface.The interlocking arm was inspected, and no anomalies was noticed.The dimensional inspection revealed that the outer diameter (od) at the zap tip and primary coil were found to be within specification.However, the number of fiber bundles were missing, due to the stretched coil condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13877902
Report Number2134265-2022-02905
Device Sequence Number14582753
Product Code KRD
UDI-Device Identifier08714729845324
UDI-Public08714729845324
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/18/2024
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0027517190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
Patient Weight60 KG
-
-