The manufacturer previously reported incorrect information alleging an issue related to the cpap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.Section b5 should have been reported as: a cpap was returned to the manufacturer for service.There was no patient harm or injury reported.During the evaluation at the manufacturer's service center, customer's complaint could not be duplicated.The top enclosure, warning label, flow manifold bottom enclosure, inlet seal, foam, blower kit, housing, cap, humidifier cable, right side assembly and outside painted panel were replaced to address the issue.Section e.1 was incorrectly reported in previous report, which is now updated in this follow-up report.
|