Model Number 561CA
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2014
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to the cpap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Additional Manufacturer Narrative
The manufacturer previously reported incorrect information alleging an issue related to the cpap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res (b.Section b5 should have been reported as: a cpap was returned for service with an allegation that the unit would no power on.There was no patient harm or injury reported.The device was evaluated by the manufacturer and a visual inspection of the remstar found evidence of dark-colored liquid contaminant on the top enclosure.Removal of the top enclosure overlay also found evidence of similar contaminants.The internal inspection of the remstar found evidence of a thermal failure to pca components (display side) j2, c29, u1, cr2, q8, and components proximal to these components.Additionally, there is evidence of thermal damage found to the non-display side of the pca.There is evidence of a dark-colored liquid on the pca (similar to that noted previously).The reported event is a result of the remstar pca being exposed to an unknown dark-colored liquid contaminant.Section e.1 was incorrectly reported in previous report, which is now updated in this follow-up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR AUTO A-FLEX
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13878146
Report Number2518422-2022-11023
Device Sequence Number17316420
Product Code BZD
UDI-Device Identifier00606959030527
UDI-Public00606959030527
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number561CA
Device Catalogue Number561CA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2014
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2014
Supplement Date Received by Manufacturer07/21/2014
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/15/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Patient SexUnknown
-
-