MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 7MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 7MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036127200
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code Vasoconstriction (2126)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This is 1 of 2 reports (1st mdr).The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during coil embolization of an aneurysm at the bifurcation of the ophthalmic artery, sl-10 and headway 17 microcatheters were placed in the aneurysm, and atlas stent was placed for coil embolization using the jail method.When the coil (subject device) was inserted as the first coil, the sl-10 microcatheter was kicked back out of the aneurysm.While retracting the coil (subject device), there was a feeling of being stuck in the atlas due to strong resistance.When the coil (subject device) was slowly retracted, the detachment part was detached.Thus, the microcatheter was retracted slowly, the coil (subject device) has also come along with the microcatheter, then it happened to drag and migrate the atlas stent.A gooseneck used as a medical intervention to retrieve the coil (subject device).At that time, because the coil (subject device) was entangled in the strut of the atlas, the atlas was retrieved together with the coil (subject device), and the coil (subject device) was removed in the guide catheter (optimo 9fr).There was some spasm, which may have been caused by the stimulation of stent removal, but it resolved with time.A ct scan was done immediately after the procedure and showed no bleeding.The procedure was completed successfully.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added h3 device evaluated by mfg ¿updated h3 summary attached - updated d4 expiration date - added d9 product available to stryker ¿ updated d9 returned to manufacturer on ¿updated.Due to the automated mes (manufacturing execution system) system, there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the coil delivery wire was found to be kinked/bent, the coil delivery wire was found to be physically detached from the main coil and the main coil was not returned.A functional test for the reported event main coil prematurely detached/separated inside patient was confirmed as the delivery wire was confirmed to be physically detached from the main coil.A functional test for device interaction with another device, main coil tangled, patient vasospasm non-serious, and main coil stretched was not available.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported 'main coil prematurely detached/separated during use', was confirmed during analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.It was reported that the subject coil was inserted as the first coil, but the microcatheter was kicked back out of the aneurysm, so the subject coil was retracted.While retracting the subject coil, there was feeling of stuck in the stent due to strong resistance.Since the microcatheter could not be reinserted, when the subject coil was slowly retracted, the detachment part was detached.Thus, the microcatheter was retracted slowly, the subject coil was also come along with the microcatheter, then it happened to drag and migrate the stent.At the result, the subject coil was completely out of the microcatheter (left inside of the body).Then used a gooseneck snare to retrieve the coil.Analysis of the returned device revealed the coil delivery wire was kinked/bent and the main coil had been prematurely detached/separated from the delivery wire, therefore, the as reported ¿main coil prematurely detached/separated during use' was confirmed.The reported 'device interaction with another device' and 'patient vasospasm non-serious' could not be confirmed as the events are patient/procedure related.The reported 'main coil tangled' and 'main coil stretched' could not be confirmed as the main coil was not returned for analysis.The analysis results are consistent with the reported events.It is probable that the main coil microcatheter kick back caused the main coil to become tangled within the struts of the stent, and the main coil was subsequently stretched and prematurely detached during the attempt to remove the microcatheter and coil.An assignable cause of caused by other will be assigned to the reported and analyzed ¿main coil prematurely detached/separated during use' and to the reported 'device interaction with another device', 'main coil tangled'.'patient vasospasm non-serious', and 'main coil stretched¿ as the investigation indicates another product(s) caused the issue event.It is probable that the delivery wire proximal end was kinked due to handling of the device during use, therefore an assignable cause of handling damage will be assigned to the analyzed 'coil delivery wire kinked/bent'.
 
Event or Problem Description
It was reported that during coil embolization of an aneurysm at the bifurcation of the ophthalmic artery, sl-10 and headway 17 microcatheters were placed in the aneurysm, and atlas stent was placed for coil embolization using the jail method.When the coil (subject device) was inserted as the first coil, the sl-10 microcatheter was kicked back out of the aneurysm.While retracting the coil (subject device), there was a feeling of being stuck in the atlas due to strong resistance.When the coil (subject device) was slowly retracted, the detachment part was detached.Thus, the microcatheter was retracted slowly, the coil (subject device) has also come along with the microcatheter, then it happened to drag and migrate the atlas stent.A gooseneck used as a medical intervention to retrieve the coil (subject device).At that time, because the coil (subject device) was entangled in the strut of the atlas, the atlas was retrieved together with the coil (subject device), and the coil (subject device) was removed in the guide catheter (optimo 9fr).There was some spasm, which may have been caused by the stimulation of stent removal, but it resolved with time.A ct scan was done immediately after the procedure and showed no bleeding.The procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 7MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13879095
Report Number3008881809-2022-00136
Device Sequence Number5264307
Product Code HCG
UDI-Device Identifier07613327004632
UDI-Public07613327004632
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/17/2024
Device Model NumberM0036127200
Device Catalogue NumberM0036127200
Device Lot Number23043497
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATLAS STENT (STRYKER); HEADWAY MICROCATHETER (NON-STRYKER); OPTIMO 9FR GUIDE CATHETER (NON-STRYKER); SL-10 MICROCATHETER (STRYKER)
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-