MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TRANSFORM COMPLIANT 5MM X 10MM - US; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TRANSFORM COMPLIANT 5MM X 10MM - US; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY Back to Search Results
Model Number M003SRC05100
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during procedure the balloon (subject device) ruptured when inserted inside pipeline stent.The procedure was completed successfully without a balloon.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added d4 expiration date - added due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.It was reported that the subject balloon ruptured when inserted inside the pipeline stent during post processing.Additional information provided by the customer indicated that no anomalies were noted to the device after removal from the packaging or prior to preparation, the device was prepared as per the dfu, the balloon inflated during prep, the correct amount of inflation media was used to inflate the balloon, the device was not inflated over nominal pressure and no excessive pressure was used, 50% contrast was used during preparation and procedure, no leakage was noted during preparation, continuous flush was maintained throughout the procedure, a pipeline device was used when the issue occurred, the rupture was noticed inside the patient during use, and the procedure was completed successfully without a balloon.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint.
 
Event or Problem Description
It was reported that during procedure the balloon (subject device) ruptured when inserted inside pipeline stent.The procedure was completed successfully without a balloon.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSFORM COMPLIANT 5MM X 10MM - US
Common Device Name
CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13884915
Report Number3008881809-2022-00141
Device Sequence Number19080393
Product Code MJN
UDI-Device Identifier07613327004687
UDI-Public07613327004687
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K122576
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2022
Device Model NumberM003SRC05100
Device Catalogue NumberM003SRC05100
Device Lot Number22644193
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer05/23/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PIPELINE DEVICE
Patient SexUnknown
-
-