• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; DOUBLE FENESTRATED GRASPER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; DOUBLE FENESTRATED GRASPER Back to Search Results
Model Number 420189-11
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a da vinci-assisted pulmonary segmentectomy surgical procedure, the robotic instrument tip articulated and dangled.Surgeon had to cut the tip off of it in order to remove it out of the patient.Instrument fragments reportedly fell inside the patient, but all pieces were retrieved without injury.The procedure was completed as planned with no reported injury.On 23-feb-2022, intuitive surgical, inc (isi) received user facility report (b)(4) stating: "robotic instrument tip articulated and dangled.Surgeon had to cut the tip off of it in order to remove it out of the patient.All pieces were retrieved without injury." on (b)(6) 2022, isi spoke to the risk manager and obtained the following information: the part number was 420189-11 and it was for a double fenestrated grasper instrument.She did not know where the instrument was currently, but will look around to see if it is in the risk management office (she returns in the office later this week).There was no mention of the instrument colliding with another instrument or with a hard object inside the patient.The surgeon was articulating the instrument when it broke.She did not know if the instrument was inspected before use, because she was not there.She was unsure how long the instrument was in use before the instrument broke.The fragment was retrieved in the same procedure, but she did not know what type of instrument retrieved the fragment.There was no post operative tests performed to check for remaining fragments.The procedure was completed robotically.On 09-mar-2022 the risk manager stated she checked with the operating room and materials management and they do not have the instrument or fragment so she is unable to return the items back to isi.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) has not yet received the double fenestrated grasper instrument for evaluation, therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the instrument is returned (post failure analysis evaluation) or if additional information is received.A review of the instrument log of the double fenestrated grasper (part # 420189-11; / lot # n10190128-547) associated with this event has been performed.Per the logs, the instrument was last used on (b)(6) 2021 on system sh1182.The alleged event occurred on the 9th use of the instrument.A review of the site's complaint history does not show any additional complaints for this product.No image or procedure video was provided for review.This complaint is being reported due to the following conclusion: it was alleged during a da vinci-assisted pulmonary segmentectomy surgical procedure, the surgeon had to cut the damaged tip off a double fenestrated grasper instrument to remove it from the patient.Instrument fragments reportedly fell inside the patient, and all fragments were retrieved successfully during the same procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
DOUBLE FENESTRATED GRASPER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key13887598
Report Number2955842-2022-10749
Device Sequence Number991390
Product Code NAY
UDI-Device Identifier00886874111581
UDI-Public(01)00886874111581(10)N10190128
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Risk Manager
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420189-11
Device Catalogue Number420189
Device Lot NumberN10190128 547
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient Age70 YR
Patient SexFemale
Patient Weight74 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-