MEDACTA INTERNATIONAL SA GMK-REVISION FIXED TIBIAL INSERT SC SIZE 4/14MM Back to Search Results
Model Number 02.07.0414SCF
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 02 march 2022: lot 2003167: (b)(4) items manufactured and released on 22-july-2020.Expiration date: 2025-07-09.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Clinical evaluation performed by medical affairs department: delayed infection in tka, 4 months after surgical operation.Infection is a known possible adverse event following every surgery, including tka's.To date, there is no reason to suspect that the cause may be linked to the implanted devices.Preliminary investigation performed by r&d knee manager: the pictures show the tibial insert explanted covered by biological fluids, as expected after implant removal.Based on the available information it is not possible to state an implant's related failure root cause.
 
Event or Problem Description
Revision surgery 4 months after primary for knee infection.Liner revised successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-REVISION FIXED TIBIAL INSERT SC SIZE 4/14MM
Common Device Name
FIXED TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
CH  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key13888076
Report Number3005180920-2022-00197
Device Sequence Number16494243
Product Code JWH
UDI-Device Identifier07630030817878
UDI-Public07630030817878
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K103170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/09/2025
Device Model Number02.07.0414SCF
Device Catalogue Number02.07.0414SCF
Device Lot Number2003167
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-