ZIMMER SWITZERLAND MANUFACTURING GMBH BIOLOX DELTA, CERAMIC FEMORAL HEAD, L, ø 36/+3.5, TAPER 12/14; BIOLOX DELTA CERAMIC FEMORAL HEADS, BIOLOX OPTION CERAMIC FEMORAL HEAD SYSTEM Back to Search Results
Model Number N/A
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Unintended Movement (3026)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 11/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Zimmer reference: (b)(4).
 
Event or Problem Description
It was reported that the patient underwent a revision due to dislocation.
 
Additional Manufacturer Narrative
Investigation results were made available.As no product was returned, visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Device is used for treatment.Insufficient information provided.Unable to perform a compatibility check.Review of complaint history identified no additional similar complaints for the reported item and no additional complaints for the reported part and lot combination related to the event.Keyword search : dislocated.Medical records were not provided.The investigation did not identify a non-conformance or a complaint out of box (coob).It has been confirmed that the implant is not within the scope or subject of any field actions or recalls which could be attributed to reported event.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The need for corrective measures is not indicated and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet's reference number of this file is (b)(4).
 
Event or Problem Description
No event update.Investigation results are now available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA, CERAMIC FEMORAL HEAD, L, ø 36/+3.5, TAPER 12/14
Common Device Name
BIOLOX DELTA CERAMIC FEMORAL HEADS, BIOLOX OPTION CERAMIC FEMORAL HEAD SYSTEM
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13888569
Report Number0009613350-2022-00180
Device Sequence Number990507
Product Code LZO
UDI-Device Identifier00889024430372
UDI-Public00889024430372
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K192416
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00-8775-036-03
Device Lot Number3069802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Supplement Date Received by Manufacturer06/20/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age47 YR
Patient SexFemale
Patient Weight78 KG
-
-