• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNITED STATES ENDOSCOPY GROUP, INC. RAPTOR GRASPING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNITED STATES ENDOSCOPY GROUP, INC. RAPTOR GRASPING DEVICE Back to Search Results
Model Number 00711177
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that during a patient procedure that included use of a raptor grasping device their forceps would not grasp a feeding tube without sliding resulting in a procedure delay.No report of patient harm.
 
Additional Manufacturer Narrative
The device subject of the event was not returned for evaluation.Without the return of the device for evaluation, the root cause of the event could not be determined.The lot number subject of the event in unknown.The raptor grasping device instructions for use states, "actuate the device by moving the slider on the handle back and forth to confirm that the grasping jaws open and close smoothly.If the unit does not function properly, or there is evidence of damage (e.G.Bends, kinks, misshapen jaws, misaligned jaws, exposed wires) do not use this product and contact your local product specialist.Do not use excessive force on the handle and do not coil the catheter outside of the endoscope.Excessive force or coiling may damage the device or damage the endoscope and may result in accidental injury to the patient or clinician.The following conditions may cause the device to function improperly: advancing the handle to the open position with too much speed or force.Attempting to pass or open the device in an extremely articulated endoscope.Attempting to actuate the device in an extremely coiled position.Actuating the device when the handle is at an acute angle in relation to the sheath." no additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPTOR GRASPING DEVICE
Common Device Name
GRASPING DEVICE
Manufacturer (Section D)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer (Section G)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer Contact
coletta cohara
5976 heisley road
mentor, OH 44060
4403586251
MDR Report Key13889139
Report Number1528319-2022-00014
Device Sequence Number15034741
Product Code OCZ
UDI-Device Identifier00724995183653
UDI-Public(01)00724995183653
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00711177
Device Catalogue Number00711177
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-