Model Number 97702
Device Problems Failure to Deliver Energy (1211); Energy Output Problem (1431); Loss of Data (2903)
Patient Problems Device Overstimulation of Tissue (1991); Discomfort (2330)
Event Date 03/20/2022
Event Type  malfunction  
Event Description
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported that the patient stated that she went up to go to the restroom and it felt like stimulator went "full blast".Patient typically runs at 1.3 v and does not touch settings.When using pp to check ins, the  therapy was off and amplitude was set to 0.Before calling into technical services (ts), the patient was warned that she may feel stimulation during an impedance check.Patient felt uncomfortable stim from neck to toes and across lumbar on back during impedances.Patient's leads are placed cervically.Impedances are all within normal range.Patient has not touched settings since device was implanted.The issue was not resolved through troubleshooting.Ts advised that there is a chance that lead could have migrated.Suggested to having imaging done and compare them to older images if possible to determine if migration is possible.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information from the rep indicated that the cause of the issue was unknown.The patient was to meet with the hcp on (b)(6) 2022.Impedances were checked, and the rep made sure the device was off.The issue was not resolved.The hcp was going to to some tests and x-rays.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13889768
MDR Text Key287824284
Report Number3004209178-2022-03698
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169974883
UDI-Public00643169974883
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2020
Device Model Number97702
Device Catalogue Number97702
Was Device Available for Evaluation? No
Date Manufacturer Received03/31/2022
Date Device Manufactured03/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
Patient Weight62 KG
-
-