• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDICREA INTERNATIONAL SA UNID; ORTHOSIS, SPINAL PEDICLE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICREA INTERNATIONAL SA UNID; ORTHOSIS, SPINAL PEDICLE FIXATION Back to Search Results
Model Number B33126012
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from post market clinical study and healthcare provider via manufacturer representative regarding an event happened during post-op of the reported products.The pre-op diagnosis was neurological spine deformity and medical history includes high blood pressure, smoker, posterior arthrodesis l1s1 with osteotomy.It was reported that, there was a device problem - drop of instrumentation.It was unknown, if there any additional treatment or surgery performed as a result.  there was no patient symptoms reported.No further complications reported regarding the event.
 
Event or Problem Description
Information was received from post market clinical study and healthcare provider via manufacturer representative regarding an event happened during post-op of the reported products.The pre-op diagnosis was neurological spine deformity and medical history includes high blood pressure, smoker, posterior arthrodesis l1s1 with osteotomy.It was reported that, there was a device problem - drop of instrumentation.It was unknown, if there any additional treatment or surgery performed as a result.Additional information received.It was reported that, vertebrae at the top/ pjk (proximal junction kyphosis).There was no malfunction associated with reported products.There was a revision surgery assembly extension performed on 11/jan/2017.No further complications reported.Mdr decision corrected to not reportable.No additional supplemental required unless additional information received indicates reportable event.
 
Additional Manufacturer Narrative
Review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNID
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION
Manufacturer (Section D)
MEDICREA INTERNATIONAL SA
5389 rte de strasbourg
rillieux la pape,fr 69140
FR  69140
Manufacturer (Section G)
MEDICREA INTERNATIONAL SA
5389 rte de strasbourg
rillieux la pape,fr 69140
FR   69140
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key13891498
Report Number1000432246-2022-00019
Device Sequence Number1779941
Product Code MNI
UDI-Device Identifier03613720205760
UDI-Public03613720205760
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K140738
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB33126012
Device Catalogue NumberB33126012
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date07/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight63 KG
-
-