• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD UD 27IN 3-0 S/A CP-2; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD UD 27IN 3-0 S/A CP-2; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J868H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.What is the specific number of devices (total quantity involved) that failed during this procedure? what tissue was being approximated or sutured when it broke? what was used to complete the procedure? procedure name and date? event date? a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported via: 2210968-2022-02138.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure in 2022 and suture was used.During the procedure, the suture broke.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 4/26/2022 additional information: d9, h3, h6 h6 component code: g07002 - no device problem found h3 investigational summary: one opened box with eight packets of product code j868h was returned to ethicon inc for analysis with the packaging closed.Upon initial inspection of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned sample, the eight packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to damaged or breakage and no defects were observed during evaluation.A functional test was performed using an instron and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD UD 27IN 3-0 S/A CP-2
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13891816
Report Number2210968-2022-02139
Device Sequence Number11191105
Product Code GAM
UDI-Device Identifier10705031043329
UDI-Public10705031043329
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberJ868H
Device Catalogue NumberJ868H
Device Lot NumberLEM041
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-