• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KINETIC CONCEPTS, INC. V.A.C.® DRAPE; OMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KINETIC CONCEPTS, INC. V.A.C.® DRAPE; OMP Back to Search Results
Model Number VACDSP
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Fungal Infection (2419)
Date of Event 02/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluated my mfr: the v.A.C.® drape was not returned; therefore, a device evaluation could not be performed.Based on the information provided, it cannot be determined that the alleged yeast infection is related to the v.A.C.® drape.All end release testing of the product and packaging met specifications.Device labeling, available in print and online, states: infected wounds should be monitored closely and may require more frequent dressing changes than non-infected wounds, dependent upon factors such as wound conditions and treatment goals.Refer to dressing application instructions (found in v.A.C.® dressing cartons) for details regarding dressing change frequency.As with any wound treatment, clinicians and patients/caregivers should frequently monitor the patient's wound, periwound tissue and exudate for signs of infection, worsening infection or other complications.Some signs of infection are fever, tenderness, redness, swelling, itching, rash, increased warmth in the wound or periwound area, purulent discharge or strong odor.Infection can be serious, and can lead to complications such as pain, discomfort, fever, gangrene, toxic shock, septic shock and/or fatal injury.Some signs or complications of systemic infection are nausea, vomiting, diarrhea, headache, dizziness, fainting, sore throat with swelling of the mucus membranes, disorientation, high fever, refractory and/or orthostatic hypotension or erythroderma (a sunburn-like rash).If there are any signs of the onset of systemic infection or advancing infection at the wound site, contact the treating physician immediately to determine if v.A.C.® therapy should be discontinued.Precautions the v.A.C.® therapy system will not be effective in addressing complications associated with the following: ischemia to the incision or incision area.Untreated or inadequately treated infection.Inadequate hemostasis of the incision.Cellulitis of the incision area.
 
Event or Problem Description
On 25-feb-2022, the following information was reported to kci by the patient: on (b)(6) 2022, the activ.A.C.¿ ion progress¿ remote therapy monitoring system was removed allegedly due to skin irritation, pain and patient developed a fungus around the wound due to moisture.Patient using a fungal powder and had a follow up with the vascular surgeon on (b)(6) 2022.The activ.A.C.¿ ion progress¿ remote therapy monitoring system will be on hold for 2 weeks.On (b)(6) 2022, the following information was reported to kci by the nurse: on (b)(6) 2022, the nurse stated patient had some redness and irritation under the v.A.C.® drape.The physician did ordered nystatin powder to treat the fungal infection caused by moisture under the v.A.C.® drape.On (b)(6) 2022, a device history record review for v.A.C.® granufoam¿ dressing lot number 9190886v009 and associated v.A.C.® drape was completed.All end release testing of the product and packaging met specifications.
 
Event or Problem Description
On 24-feb-2022, the following information was reported to kci by the patient: on 21-feb-2022, the activ.A.C.¿ ion progress¿ remote therapy monitoring system was removed allegedly due to skin irritation, pain and patient developed a fungus around the wound due to moisture.Patient using a fungal powder and had a follow up with the vascular surgeon on (b)(6) 2022.The activ.A.C.¿ ion progress¿ remote therapy monitoring system will be on hold for 2 weeks.On 23-mar-2022, a device history record review for v.A.C.® granufoam¿ dressing lot number 9138894v009 and associated v.A.C.® drape was completed.All end release testing of the product and packaging met specifications.
 
Additional Manufacturer Narrative
Mdr-3009897021-2022-00064 submitted on 24-mar-2022 noted the following: section b5 describe event or problem: on 25-feb-2022, the following information was reported to kci by the patient: on 23-mar-2022, a device history record review for v.A.C.® granufoam¿ dressing lot number 9190886v009; section d4 lot#: 9190886v009; section d4 unique identifier (udi) #: (b)(4).Section b5 describe event or problem: on 24-feb-2022, the following information was reported to kci by the patient: on 23-mar-2022, a device history record review for v.A.C.® granufoam¿ dressing lot number 9138894v009; section d4 lot#: 9138894v009; section d4 unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V.A.C.® DRAPE
Common Device Name
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
san antonio TX 78249
Manufacturer Contact
steven jackson
6203 farinon drive
san antonio, TX 78249
2102556438
MDR Report Key13891953
Report Number3009897021-2022-00064
Device Sequence Number12325110
Product Code OMP
UDI-Device Identifier00849554002098
UDI-Public010084955400209817240331109138894V009
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K133276
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberVACDSP
Device Catalogue NumberM8275052
Device Lot Number9138894V009
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer02/24/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACETAMINOPHEN 325MG TAB 2 TABS PO EVERY 6 HRS PRN.; ASPIRIN 325MG EC TAB, 1 TAB PO DAILY.; CLOPIDOGREL 75MG TAB, 1 TAB PO DAILY.; DOCUSATE SODIUM 100MG CAPSULE 1 CAPSULE EVERY MORN.; DULAGLUTIDE 3MG/0.5ML INJECTION, INJECT 0.5ML UNDE.; EVOLOCUMAB 140MG/ML SYRINGE, INJECT 140MG UNDER TH.; FAMOTIDINE 20MG TAB, 1 TAB PO EVERY MORNING AND EV.; FENOFIBRATE 48MG TAB, 1 TAB EVERY MORNING AND EVER.; IBUPROFEN 600MG TABS, 1 TAB PO EVERY 6 HRS AS NEED.; LISINOPRIL 10MG TAB, 1 TAB PO DAILY.; MELATONIN 10MG CAPSULE, 1 CAPSULE PO AT BEDTIME AS.; METFORMIN 1000MG TAB, 1 TAB PO WITH BREAKFAST AND.; METOPROLOL ER 25MG 24HR TAB, 1 TAB PO DAILY.; MULTIVITAMIN CAPSULE, 1 CAPSULE PO DAILY.; OMEGA 3 FATTY ACIDS/ FISH OIL 300-1,000MG CAPSULE.; PSYLLIUM 3.4 GRAM PACKET, 1 PACKET PO DAILY.; SILDENAFIL 20MG TAB, 5 TABS 1 HR PRIOR TO SEXUAL A.; TRAZODONE 100MG TAB, 1 TAB PO AT BEDTIME FOR 90 DA.; TRIAMCINOLONE 0.1% OINTMENT, APPLY TOPICALLY TO EL.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexMale
-
-