• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD¿ PHLEBOTOMY SHARPS COLLECTOR, 1 QT WITH NEEDLE PORT; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD¿ PHLEBOTOMY SHARPS COLLECTOR, 1 QT WITH NEEDLE PORT; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305635
Medical Device Problem Code Missing Information (4053)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Oem manufacturer: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: na.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd¿ phlebotomy sharps collector, 1 qt with needle port there was no label or missing label information.This event occurred with 5 collectors.The following information was provided by the initial reporter.The customer stated: "there was no label on five products in a carton.".
 
Additional Manufacturer Narrative
H6: investigation summary: no photos or samples were received.Additional attempt to get more information was made, however, the customer confirmed that no additional information was available.According to the dhr review process, the result showed there were no issues reported as missing label during the manufacturing process of the lot number reported (1268908) under this customer complaint.
 
Event or Problem Description
It was reported when using the bd¿ phlebotomy sharps collector, 1 qt with needle port there was no label or missing label information.This event occurred with 5 collectors.The following information was provided by the initial reporter.The customer stated: "there was no label on five products in a carton.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ PHLEBOTOMY SHARPS COLLECTOR, 1 QT WITH NEEDLE PORT
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13892095
Report Number2243072-2022-00390
Device Sequence Number1713755
Product Code FMI
UDI-Device Identifier00382903056354
UDI-Public00382903056354
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943139
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number305635
Device Catalogue Number305635
Device Lot Number1268908
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-