• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXPANDING INNOVATIONS, INC. X-PAC; SPINAL CAGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXPANDING INNOVATIONS, INC. X-PAC; SPINAL CAGE Back to Search Results
Model Number EI2511L
Health Effect - Clinical Code Intervertebral Disc Compression or Protrusion (4524)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A complete investigation was initiated upon receipt of the complaint.The product was not returned for evaluation, so a review of specific lot production records was not possible.However, pictures provided by the user facility showed signs of migration posteriorly.It was noted that the original placement of the cage may have been more posterior than normal and that the patient had poor bone quality.It was confirmed upon removal of the cage that there was no decrease in height.The exact cause of the cage migration cannot be determined.
 
Event or Problem Description
It was reported by the distributor representative that posterior spinal fusion surgery had been performed on a female patient at l5/s1 on (b)(6) 2021.During patient follow up visit to doctor on (b)(6) 2022, it was observed that migration of the implanted cage had occurred posteriorly and that the cage was impinging on the nerve.Revision surgery was scheduled for (b)(6) 2022.On (b)(6) 2022, it was reported that the revision surgery had taken place on the scheduled date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-PAC
Common Device Name
SPINAL CAGE
Manufacturer (Section D)
EXPANDING INNOVATIONS, INC.
110pioneer way, suite i
mountian view CA 94041
Manufacturer (Section G)
EXPANDING INNOVATIONS, INC.
110pioneer way, suite i
mountian view CA 94041
Manufacturer Contact
mbithi muthini
110 pioneer way, suite i
mountain view, CA 94041
6508613129
MDR Report Key13892291
Report Number3007289746-2022-00001
Device Sequence Number6356835
Product Code MAX
UDI-Device Identifier00850001720298
UDI-Public00850001720298
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K201145
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEI2511L
Device Catalogue NumberEI2511L
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-