ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MV-13-T; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 133S-MV-13-T
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
Healthcare professional reported that a tissue expander "would not hold fluids." the device was never implanted.It is unknown if a replacement device was used to complete surgery.
 
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: tissue expander "would not hold fluids.".
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.A visual analysis of the returned device identified: crease flat, red particles outer device and white particles outer device.Leak test and microscopic analysis was performed which identified: device no leak.Based on the device analysis the final assessment is: there was no opening and leak was observed in the device.
 
Event or Problem Description
Healthcare professional reported that a tissue expander "would not hold fluids." the device was never implanted.It is unknown if a replacement device was used to complete surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATRELLE TE SMOOTH 133S-MV-13-T
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key13892299
Report Number9617229-2022-04044
Device Sequence Number12351610
Product Code LCJ
UDI-Device Identifier10888628043879
UDI-Public10888628043879
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K182054
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/11/2025
Device Model Number133S-MV-13-T
Device Catalogue Number133S-MV-13-T
Device Lot Number3456061
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer03/30/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-