| Model Number |
133S-MV-13-T |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Healthcare professional reported that a tissue expander "would not hold fluids." the device was never implanted.It is unknown if a replacement device was used to complete surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: tissue expander "would not hold fluids.".
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.A visual analysis of the returned device identified: crease flat, red particles outer device and white particles outer device.Leak test and microscopic analysis was performed which identified: device no leak.Based on the device analysis the final assessment is: there was no opening and leak was observed in the device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported that a tissue expander "would not hold fluids." the device was never implanted.It is unknown if a replacement device was used to complete surgery.
|
| |
|