| Model Number |
ADM04012013P |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting to use inpact admiral dcb along with non-medtronic 6fr x 45cm sheath and 300m guidewire during procedure to treat a little calcified lesion in the left mid superficial femoral artery (sfa).The vessel was none tortuous.The vessel diameter and lesion length are 4mm and 80mm respectively.There was no embolic protection use.A non-medtronic inflation device was used for balloon inflation.A 50/50 hearinized saline and contrast was used or balloon inflation.There was no damage noted to packaging.There was no issue noted when removing the device from hoop/tray.The device was prepper per ifu with no issues identified.It was reported that during balloon inflation, balloon burst occurred at 14atm.All fragments of the balloon were retrieved.The device passed through a previously deployed stent.There was no resistance encountered when advancing the device.There was no intervention need to remove the burst balloon.There was no vessel damage due to event.The device was safely removed from patient.The defective balloon was removed, and a new balloon was inserted and inflated to complete the procedure.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|