| Model Number |
CL-10 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Chemical Exposure (2570); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event involved a chemolock¿ bag spike which was reported to have an irinotecan bag poking through a baxter port resulting in a leak.The event happened when the nurse was ready for administration, it came into contact with the nurse, and the spill was cleaned up.The drug was remixed using chemolock components and a new bag of medication was replaced.There was a hole in the port.There was patient involvement, but no harm reported as a consequence of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available for evaluation, however, sample photo has been provided, but investigation not yet complete.
|
| |
|
Additional Manufacturer Narrative
|
|
No product samples were returned for investigation, however, a representative photograph was returned that shows the spike tip of a cl-10 bag spike that has penetrated through the sidewall of the infusion bag access port.There does not appear to be any cl-10 bag spike damage or anomalies visible from the photograph returned.The device history review was performed and there were no relevant non-conformances found that would have contributed to the reported complaint.A probable cause of the complaint cannot be determined without return of the affected sample.
|
| |
|