• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. CHEMOLOCK¿ BAG SPIKE; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. CHEMOLOCK¿ BAG SPIKE; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM Back to Search Results
Model Number CL-10
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Chemical Exposure (2570); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
The event involved a chemolock¿ bag spike which was reported to have an irinotecan bag poking through a baxter port resulting in a leak.The event happened when the nurse was ready for administration, it came into contact with the nurse, and the spill was cleaned up.The drug was remixed using chemolock components and a new bag of medication was replaced.There was a hole in the port.There was patient involvement, but no harm reported as a consequence of this event.
 
Additional Manufacturer Narrative
The device is not available for evaluation, however, sample photo has been provided, but investigation not yet complete.
 
Additional Manufacturer Narrative
No product samples were returned for investigation, however, a representative photograph was returned that shows the spike tip of a cl-10 bag spike that has penetrated through the sidewall of the infusion bag access port.There does not appear to be any cl-10 bag spike damage or anomalies visible from the photograph returned.The device history review was performed and there were no relevant non-conformances found that would have contributed to the reported complaint.A probable cause of the complaint cannot be determined without return of the affected sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHEMOLOCK¿ BAG SPIKE
Common Device Name
CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key13893142
Report Number9617594-2022-00063
Device Sequence Number6252104
Product Code ONB
UDI-Device Identifier00887709056961
UDI-Public(01)00887709056961(17)260901(10)5589706
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K131549
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2026
Device Model NumberCL-10
Device Catalogue NumberCL-10
Device Lot Number5589706
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date04/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
IRINOTECAN, MFR UNK; PORT, BAXTER
Patient SexUnknown
-
-