| Model Number |
1550300-15 |
| Medical Device Problem Code |
Expiration Date Error (2528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the radio frequency identification (rfid) consignment tracking tag for the xience sierra stent delivery system (sds) is showing the wrong expiration year, the month and date are correct.The expiration date on the rfid tag is showing 12/22/2008.There was no device use or patient involvement.There was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated the device was not returned for analysis.A review of the electronic lot history record (elhr) and a lot level similar incident review for this product was not performed because the lot number was not reported and the product was not returned for analysis.The investigation determined the reported difficulties appear to be related to the format of the accounts barcode scanning setup as the account confirmed that their expiration date layout is set up differently than abbott vascular as they use dd/mm/yyyy format and abbott vascular barcode dates are set up yy/mm/dd.There is no indication of a product quality issue with respect to manufacture, design or labeling.The additional xience sierra device is being filed under a separate medwatch report number.
|
| |
|