Model Number 7005
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The device was visually and microscopically examined.There was a hypotube kink at 56cm from the strain relief.There was a separation in the hypotube at 87.5cm from the strain relief.The separated ends of the hypotube were oval shaped indicating that the device was kinked prior to separation.There was contrast in the inflation lumen.The rapid port exchange was damaged, and the balloon was tightly folded with blood and contrast present in the balloon folds.Photo reviewed depicted separation to the device which aligns to the separation found in product analysis.The damage to the device is consistent to what can occur during the procedure or an interaction with patient characteristics.
 
Event or Problem Description
Reportable based on device analysis completed on 01-mar-2022.It was reported that the device was damaged.A 3.5mm x 8mm quantum maverick balloon catheter was selected for use.However, after unpacking, it was found that the device was damaged and could not be used.No patient complications were reported and the patient status was stable.However, device analysis revealed a hypotube separation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13895242
Report Number2134265-2022-02395
Device Sequence Number4470013
Product Code LOX
UDI-Device Identifier08714729392583
UDI-Public08714729392583
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/23/2023
Device Model Number7005
Device Catalogue Number7005
Device Lot Number0026063225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-