Model Number 24681
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified vein in the left upper arm a 8.0 x40, 75cm charger balloon catheter was advanced for dilatation, however, during first inflation at 18 atmospheres, the balloon ruptured and balloon leak was noted.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
A charger device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 22mm in length and extended from a position 3mm distal of the proximal markerband to 25mm distal of the proximal markerband.A visual examination observed damage to the tip of the device.A visual and tactile examination of the shaft of the device found no issues.A visual examination found no issues or damage to the markerbands of the device.This concludes the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified vein in the left upper arm a 8.0 x40, 75cm charger balloon catheter was advanced for dilatation, however, during first inflation at 18 atmospheres, the balloon ruptured and balloon leak was noted.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARGER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13895296
Report Number2134265-2022-03409
Device Sequence Number4303626
Product Code FGE
UDI-Device Identifier08714729804673
UDI-Public08714729804673
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K112697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/27/2024
Device Model Number24681
Device Catalogue Number24681
Device Lot Number0027744246
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer04/15/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
Patient Weight84 KG
Patient RaceBlack Or African American
-
-