| Model Number |
E-Series |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that one of the lightheads detached from the spring arm of their harmonyair e-series surgical lighting system.No report of injury or procedure delay.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the lighting system and found that the snap ring had become dislodged from where installed and the stop pin was broken.Due to the damage, the technician made the necessary repairs, tested the lighting system, confirmed it to be operating according to specification, and returned it to service.The observed damage to the pin is indicative of the lighthead being subject to a high load of impact by user facility personnel.The technician counseled the user facility on the proper use and operation of the harmonyair e-series surgical lighting system.No additional issues have been reported.
|
| |
|