| Model Number |
FX609T |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
Hydrocephalus (3272)
|
| Date of Event |
02/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation: visual inspection: no significant deformations or damage of the valves were detected during the visual inspection.Permeability test: a permeability test has shown that the shunt system is permeable.Computer controlled test: to investigate the claim of a blockage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operate within the accepted tolerance in both positions.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, deposits were found in both valves.Results: it should be noted that the product was received dry (i.E.Not submersed in liquid as recommended).The investigation of dry items is not significant due to the affect dry deposits of liquor and blood can have on product performance.In spite of this, we have investigated the valve to the best of our abilities.Based on our examination results, we can detect deposits in the shunt system at the time of our examination.The deposits do not affect the technical characteristics at the time of the examination.Proteins / deposits in the csf can temporarily impair the function of the valves and are described side effects in hydrocephalus therapy.Based on our investigation, we can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
| |
|
Event or Problem Description
|
|
It was reported that a progav 2.0 shunt system (part # fx609t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be clogged.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6), height: 120 centimeters (cm), weight: (b)(6), gender: unknown.
|
| |
|
Search Alerts/Recalls
|