• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER LITE 10X10CM; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER LITE 10X10CM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 66020408
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event or Problem Description
It was reported that, during treatment, when a sterile pouch of an allevyn gentle border lite 10x10cm was taken out, it was noticed that there was an 8 cm size unsealed part on one side edge.Treatment was completed with a sn back-up dressing.No harm to the patient or further complications reported.
 
Additional Manufacturer Narrative
Section h3, h6: the device was not used in treatment, has been returned and evaluated, establishing a relationship between the device and the reported event.Visual inspection observed the dressing was curled and captured at the edge of the seal.The root cause has been determined as a quality control deficiency.The dressing skewed after passing camera detection and then trapped in the side seal of pouch.The trapped dressing was not captured by operator during visual inspection.The batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.The complaint history review confirms previous complaints of this nature with manufacturing problems observed.A risk management review concluded that the alleged failure mode and any associated harm has been anticipated within the risk file, with no updates required.A review of prior escalation actions found no actions applicable to the scope of this case.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN GENTLE BORDER LITE 10X10CM
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13896538
Report Number8043484-2022-00112
Device Sequence Number4240472
Product Code NAD
UDI-Device Identifier05000223475499
UDI-Public05000223475499
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2024
Device Model Number66020408
Device Catalogue Number66800835
Device Lot Number202128
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date03/24/2022
Supplement Report FDA Received Date08/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-