• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ANGIOJET SOLENT DISTA; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ANGIOJET SOLENT DISTA; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 45030
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on the device analysis on 02-mar-2022.It was reported that an error message occurred.An angiojet solent dista was used in a thrombectomy procedure.During the procedure, it was noted that an error message was displayed.The procedure was completed with a different device.There were no patient complications reported.However, returned device analysis revealed a hypotube break.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device returned to boston scientific for analysis.The product return contained an angiojet solent dista.The assembly, supply line, shaft, tip, and spike line were all visually examined for potential damage.The device showed no signs of damage.A functional test was performed.The pump was placed into the ultra drive console and the device displayed an under pressure error message.The device was unable to run, due to the pressure being below specifications.An x-ray inspection on the device revealed a break in the hyptube at 25cm from the tip.Functional testing confirms the presence of error messages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SOLENT DISTA
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13897200
Report Number2134265-2022-03315
Device Sequence Number17159950
Product Code DXE
UDI-Device Identifier08714729889731
UDI-Public08714729889731
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K113363
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/12/2022
Device Model Number45030
Device Catalogue Number45030
Device Lot Number0026376657
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2022
Initial Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-