• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1¿, 63 CM LENGTH, 8.5F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1¿, 63 CM LENGTH, 8.5F; INTRODUCER, CATHETER Back to Search Results
Model Number 407453
Device Problem Flaked (1246)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2021
Event Type  malfunction  
Manufacturer Narrative
One 8.5f swartz braided introducer sheath and dilator were received for evaluation.The dilator and sheath were flushed with fluid and the dilator was inserted into the sheath.Resistance was noted near the curve of the dilator when and a brk needle/stylet assembly could not reach the distal end of the dilator.The dilator tubing was cut lengthwise and a build-up of skived material was noted at the distal end of the dilator.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the skiving remains unknown.
 
Event Description
This report is to advise of an event noted during analysis which revealed skiving of the dilator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1¿, 63 CM LENGTH, 8.5F
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13897236
MDR Text Key287901194
Report Number3005334138-2022-00235
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734205771
UDI-Public05414734205771
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number407453
Device Catalogue Number407453
Device Lot Number8127222
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-