AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFW225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Flashers (1864); Visual Disturbances (2140); Halo (2227)
Date of Event 01/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the intraocular lens (iol) was implanted in the patient¿s ocular sinister (left eye).The iol was explanted in a separate surgical procedure due to complaints of glare/halos and beams/rods of light at night and replaced with a different model lens with a different diopter size, diu225 18.5 diopter iol.There was no complications and no vitrectomy.Patient outcome post lens exchange was reported as: 1-day post-op stable, patient returning in three (03) weeks for post operative refraction/checkup.No further information is available.
 
Additional Manufacturer Narrative
Additional information: patient weight and ethnicity and race: unknown as information was asked but not provided.The device was not returned for analysis therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.An attempt has been made to obtain missing information; however, to date, the information has not been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation: yes section d-9: date returned to manufacturer: apr 07-2022 section h-3: device evaluated by manufacturer: yes device evaluation: the complaint lens was received stuck to a piece of paper.Visual inspection under magnification was performed and revealed that the lens was cut in half.The lens was cleaned and visually inspected under magnification and no additional issues were observed, and no issues related to or could have contributed to visual issues were observed.Due to the returned condition of the lens (cut in half) no further product evaluation can be performed.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13897301
Report Number3012236936-2022-00667
Device Sequence Number4470062
Product Code MFK
UDI-Device Identifier05050474751811
UDI-Public(01)05050474751811(17)240412
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/12/2024
Device Model NumberDFW225
Device Catalogue NumberDFW225U180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-