OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI; HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR Back to Search Results
Model Number WA64150A
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Oympus was informed that during a therapeutic laparoscopic distal pancreatectomy procedure one of the jaws at the distal end of the grasping forceps broke off and fell into the patient.However, the fragment was retrieved immediately.The intended procedure was completed and there was no report about an adverse event or patient injury.However, a ct procedure performed several days after the original procedure seems to show a minute metallic fragment.It is unclear whether this may stem from the broken off jaws of the grasping forceps.
 
Additional Manufacturer Narrative
The suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation confirmed that one jaw at the distal end of the grasping forceps is broken off at the thinnest section.In view of the instrument¿s age, this damage is considered a fatigue fracture caused by wear and tear due to frequent use/reprocessing.There are numerous small superficial scratches on the instrument indicating frequent use of the instrument.Therefore, this event/incident was attributed to use error.Furthermore, the investigation found the pull wire of the jaw insert to be heavily bent.It is assumed that this damage was intentionally induced by staff to show that this jaw insert was defective.Based on the investigation results, it can be assumed that all retrieved parts from the broken jaw insert complement each other and a complete jaw can be reconstructed with the available fragments.Accordingly, it is not apparent that a fragment from the jaw is missing.Furthermore, the typical fracture behavior of the affected material is characterized by a clean break rather than by creating many fragments.Accordingly, the probability of any remaining fragments in the patient body is assessed as very low.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the grasping forceps without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI
Common Device Name
HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key13898522
Report Number9610773-2022-00105
Device Sequence Number17260715
Product Code GCJ
UDI-Device Identifier04042761068970
UDI-Public04042761068970
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K944201
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberWA64150A
Device Catalogue NumberWA64150A
Device Lot Number15XW
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer05/19/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-