• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONSOLIDATED MEDICAL EQUIPMENT COMPANY GOLDVAC INTEGRATED SMOKE PENCIL, PUSHBUTTON (10/CS); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONSOLIDATED MEDICAL EQUIPMENT COMPANY GOLDVAC INTEGRATED SMOKE PENCIL, PUSHBUTTON (10/CS); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 60-7510-005
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
The sales representative reported on behalf of the customer that the 60-7510-005, goldvac integrated smoke pencil, pushbutton (10/cs), was used during an unknown type of procedure on (b)(6) 2022 when it was reported, ¿while finishing surgery in or-1 the cautery turned on by itself onto cut.It would not turn off till the machine was turned off.¿ there was no report of injury, medical intervention, or hospitalization for the user.The procedure was completed as planned without any delay.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The sales representative reported on behalf of the customer that the 60-7510-005, goldvac integrated smoke pencil, pushbutton (10/cs), was used during an unknown type of procedure on (b)(6) 2022 when it was reported, ¿while finishing surgery in or-1 the cautery turned on by itself onto cut.It would not turn off till the machine was turned off.¿ there was no report of injury, medical intervention, or hospitalization for the user.The procedure was completed as planned without any delay.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The device will not be returned for evaluation and no photographic evidence has been provided; therefore, the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 5 complaints, regarding 7 devices, for this device family and failure mode.During this same time frame (b)(4) have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised to always place unused associated electrosurgical accessories in a safe insulated location such as the provided holster when not in use.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOLDVAC INTEGRATED SMOKE PENCIL, PUSHBUTTON (10/CS)
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX  31136
Manufacturer (Section G)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX   31136
Manufacturer Contact
john berga
11311 concept blvd
largo, FL 33773
7273995358
MDR Report Key13899878
Report Number3007305485-2022-00039
Device Sequence Number11780836
Product Code GEI
UDI-Device Identifier30653405056335
UDI-Public(01)30653405056335(17)230920(10)202109204
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K081634
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2023
Device Model Number60-7510-005
Device Catalogue Number60-7510-005
Device Lot Number202109204
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-