BIOMET 3I 3I T3® WITH DCD® TAPERED IMPLANT 6/5 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number BNPT6510
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
It was reported that the implant was placed and would not seat properly and was remove another in the customer's inventory was used to complete the procedure.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual evaluation of the returned product identified damaged drive feature.Excessive damage as well due to the removal process.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.The reported device was located on tooth # 14 and was place and removed on the same day.No pre-existing conditions were noted on the product experience report (per).A device history record (dhr)review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.Complainant reported that the implant was placed and would not seat properly and was remove.The reported complaint was confirmed.A definitive root cause for this complaint could not be determined.The following sections have been updated: h3: device evaluated by manufacturer: change ¿no' to 'yes' h6: evaluation codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3® WITH DCD® TAPERED IMPLANT 6/5 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key13901596
Report Number0001038806-2022-00427
Device Sequence Number11822359
Product Code DZE
UDI-Device Identifier00844868029502
UDI-Public(01)00844868029502(17)260312(10)2021021168
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2026
Device Model NumberBNPT6510
Device Catalogue NumberBNPT6510
Device Lot Number2021021168
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2022
Supplement Date Received by Manufacturer08/27/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-