| Model Number |
179712850 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Initial reporter is synthes sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2022, during spinal fusion surgery the tip of the quick connect driver broke off inside an 8x50mm screw while the surgeon was inserting the screw.The surgeon removed the head to try removal with a srb set.The screw shank was unable to be extracted from the patient.Fragments were generated.There was a surgical delay of forty-five (45) minutes.Procedure was completed successfully.Patient status is unknown.This complaint involves two (2) devices.This is report 1 of 2 for (b)(4).
|
| |
|