ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 10X140MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source (b)(6).Visual evaluation of the returned product and photos provided identified foreign debris inside the sterile packaging.The sterile packaging remains sealed.Dhr was reviewed and no discrepancies were found.The likely condition when the device left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions during manufacturing.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during warehouse inspection, hair-like debris was found inside the sterile packaging.There was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 10X140MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13903070
Report Number0001825034-2022-00691
Device Sequence Number14550746
Product Code KWL
UDI-Device Identifier00880304489684
UDI-Public(01)00880304489684(17)310519(10)7024018
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-103100
Device Lot Number7024018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Initial Report FDA Received Date03/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-