Catalog Number IIPDTUL |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that when torqueing, the implant driver got stuck in the implant.The implant was removed.Another implant and tool from the customer's inventory was used to complete the procedure.
|
|
Manufacturer Narrative
|
Zimmer biomet complaint number (b)(4).
|
|
Manufacturer Narrative
|
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual evaluation of the returned product identified signs of use.No malfunction was noticed during product evaluation that will contribute to the event.Functional testing with an in-house component to recreate the reported event verified the driver easily disengaged/release.No pre-existing conditions were noted on the product experience report (per).Complainant reported that when torquing, the implant driver got stuck in the implant.The reported complaint could not be verified.A definitive root cause for this complaint could not be determined.A device history record (dhr) review could not be performed since the lot number associated to the item was not provided.However, zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming product.A complaint history review by item number was conducted for the device dating back to 12 months.The complaint history review revealed that there are no existing non-conformances/capa/hhe/d/ie/product holds for the reported product for similar events.The following sections have been updated: d9: device available for evaluation change ¿no' to 'yes'.H3: device evaluated by manufacturer: change ¿no' to 'yes'.
|
|
Event Description
|
No further event information is available at the time of this report.
|
|
Search Alerts/Recalls
|