• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I INTERNAL CONNECTION UNIVERSAL PLACEMENT DRIVER TIP - LONG; IMPLANT DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I INTERNAL CONNECTION UNIVERSAL PLACEMENT DRIVER TIP - LONG; IMPLANT DRIVER Back to Search Results
Catalog Number IIPDTUL
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that when torqueing, the implant driver got stuck in the implant.The implant was removed.Another implant and tool from the customer's inventory was used to complete the procedure.
 
Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual evaluation of the returned product identified signs of use.No malfunction was noticed during product evaluation that will contribute to the event.Functional testing with an in-house component to recreate the reported event verified the driver easily disengaged/release.No pre-existing conditions were noted on the product experience report (per).Complainant reported that when torquing, the implant driver got stuck in the implant.The reported complaint could not be verified.A definitive root cause for this complaint could not be determined.A device history record (dhr) review could not be performed since the lot number associated to the item was not provided.However, zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming product.A complaint history review by item number was conducted for the device dating back to 12 months.The complaint history review revealed that there are no existing non-conformances/capa/hhe/d/ie/product holds for the reported product for similar events.The following sections have been updated: d9: device available for evaluation change ¿no' to 'yes'.H3: device evaluated by manufacturer: change ¿no' to 'yes'.
 
Event Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERNAL CONNECTION UNIVERSAL PLACEMENT DRIVER TIP - LONG
Type of Device
IMPLANT DRIVER
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key13903386
MDR Text Key287931743
Report Number0001038806-2022-00435
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIIPDTUL
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/19/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK LB IMPLANT
-
-