MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET 20G X 1 IN; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET 20G X 1 IN; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refv3887 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "there was an incident reported that the winged infusion set 20g 1 huber needle was cracked at the end of the tubing (at the luer).".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of luer cracks could not be confirmed as the returned products did not reveal any issues corresponding to the alleged event.X59 sealed 20 ga x 1in winged infusion sets were returned for evaluation.All of the sample¿s packaging information indicated lot: refv3887.Microscopic observation of the hubs of all of the sealed samples did not reveal any apparent deformation or non-concentric hubs.No cracks in the material were noted.A 12 ml luer lock syringe was able to be attached and secured to all of the hubs without any resistance or issues.Since no apparent issues were found on the sealed samples from lot: refv3887, the complaint could not be confirmed h3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported "there was an incident reported that the winged infusion set 20g 1" huber needle was cracked at the end of the tubing (at the luer).".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WINGED INFUSION SET 20G X 1 IN
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13904541
Report Number3006260740-2022-00967
Device Sequence Number4303854
Product Code FMI
UDI-Device Identifier00801741065811
UDI-Public(01)00801741065811
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K863606
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0604200
Device Lot NumberREFV3887
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer08/15/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-