EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PHYSIO II RING; ANNULOPLASTY RING Back to Search Results
Model Number 5200M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 02/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via the implant patient registry that a 36mm mitral ring, implanted for four (4) years and six (6) months, was explanted due to severe 4+ insufficiency.The patient presented with dizziness and occasional palpitations.The explanted device was replaced with a 25mm bioprosthetic valve.The patient was discharged home in good condition on pod #6.The implantation data card indicated that the device explant was due to deficiency of the original device.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
The causes of re-operation or percutaneous intervention for failed annuloplasty repairs are well documented in the literature.Re-operations are primarily the result of a progression of disease or technical failures and are not related to product malfunctions.Unlike prosthetic heart valves, annuloplasty rings are an adjunct to the valve repair.The device was not returned for evaluation, as device request is still ongoing.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Additional Manufacturer Narrative
H10: additional narratives updated b5 corrected data: based on the additional information obtained, this event is no longer considered reportable and this correction is being submitted.
 
Event or Problem Description
It was reported via the implant patient registry that a 36mm mitral ring, implanted for four (4) years and six (6) months, was explanted due to severe 4+ insufficiency.The patient presented with dizziness and occasional palpitations.The explanted device was replaced with a 25mm bioprosthetic valve.The patient was discharged home in good condition on pod #6.Per or finding : very redundant tissue along the posterior leaflet of the mitral valve.There was stretching of the cords to the p3 which was not present at the first operation.The p2 repair was intact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PHYSIO II RING
Common Device Name
ANNULOPLASTY RING
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
reginald santos
one edwards way
mailstop anton 6.1
irvine, CA 92614
9492502731
MDR Report Key13904711
Report Number2015691-2022-04616
Device Sequence Number14469509
Product Code KRH
UDI-Device Identifier00690103180237
UDI-Public(01)00690103180237(17)220430
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K083470
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5200M
Device Catalogue Number5200M36
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Supplement Date Received by Manufacturer05/05/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age72 YR
Patient SexFemale
-
-