Catalog Number 03.010.065
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, the patient underwent the intramedullary nail fixation surgery with the mhn for the fracture of the proximal humerus.When the surgeon inserted the locking screw through the protection sleeve, the screw head did not penetrate the protection sleeve.When the surgeon pulled out the protection sleeve to check how far the screw had gone, the locking screw and the protection sleeve came out.It was confirmed that the screw head interfered with the tip of the protection sleeve many times, and the tip of the protection sleeve was slightly squashed inward.The surgeon used another protection sleeve to insert the locking screw.Procedure was completed successfully with thirty(30) minutes of surgical delay.This report is for one (1) drillsl 8/4.2 f/03.010.063.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Lot: 8326173.Manufacturing site: hägendorf.Release to warehouse date: april 04, 2013.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that drillsl 8/4.2 f/03.010.063 was observed quashed inward (bent) at the tip of sleeve.A dimensional inspection for the drillsl 8/4.2 f/03.010.063 was unable to be performed due to post manufacturing damaged.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the drillsl 8/4.2 f/03.010.063 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: 04.2 mm drill sleeve ngn.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8.0MM/4.2MM DRILL SLEEVE 200MM
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13904852
Report Number8030965-2022-01934
Device Sequence Number4392138
Product Code FZX
UDI-Device Identifier07611819760929
UDI-Public(01)07611819760929
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.065
Device Lot Number8326173
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12.0MM/8.0MM PROTECTION SLEEVE 188MM; UNK - INSERTION INSTRUMENTS: TRAUMA; UNK - SCREWS: NAIL PROXIMAL LOCKING
Patient SexUnknown
-
-