MAUDE Adverse Event Report: COOK INC HIWIRE NITINOL HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC HIWIRE NITINOL HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G30475
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, upon opening the packaging of a hiwire nitinol hydrophilic wire guide, the coating was damaged approximately 5 centimeters along the device.A photo provided shows the coating peeled/separated from the wire in several locations.This occurred prior to patient contact; there was no impact to the patient.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Event description: as reported, upon opening the packaging of a hiwire nitinol hydrophilic wire guide, the coating was damaged approximately 5 centimeters along the device.A photo provided shows the coating peeled/separated from the wire in several locations.A new device was used to complete the procedure.This occurred prior to patient contact; there was no impact to the patient.Investigation ¿ evaluation: a visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record (dhr), manufacturing instructions, the instructions for use (ifu), and quality control data.One hiwire nitinol hydrophilic wire guide was returned for investigation.The device was returned in an open package with the label, confirming the lot number.Under magnification, it appeared the coating on the wire guide has been scraped the wire guide is assembled and packaged by a cook supplier who completed an investigation as well as cook.The supplier and cook reviewed their device history records (dhr) and found no related anomalies.The incoming inspection records at cook were reviewed and the lot passed incoming inspection.A review of complaint history records shows no other complaints associated with the complaint device lot.A review of manufacturing procedures by the supplier found controls to be in place to assure device integrity prior to shipment.The evidence from the complaint file, device history record, complaint history and manufacturing documents indicates that the complaint device was manufactured to specification as well as other items in the lot or similar devices in the field or in house.Supplier investigation the supplier carried out an analysis of the returned product and noted the following the specimen presented skive/cut damage in a proximal to distal orientation located 64.8 to 64.9cm from the distal tip, exposing the metallic core wire.The specimen also presented multiple, multi-directional kinks/bends of varying severity and frequency scattered from 6.10 to 130.5cm from the distal tip.The supplier noted the nature of the skive/cut damage to the polymer jacket material appeared consistent with manipulation of the device within a metal cannula or needle and the nature of the bend damage appeared consistent with manipulation of the device against resistance.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precautions · manipulation of wire guide requires appropriate imaging control.Use caution not to force or over manipulate the wire guides when gaining access.· when using wire guide through a metal cannula/needle, use caution as damage may occur to outer coating.Based on the returned device, it appears handling and/or device interface factors have contributed to the event.However, the customer reported finding the damage when they opened the package.Therefore, cook has concluded a cause cannot be established.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Information was available but inadvertently omitted from the previous medwatch report: a new device was used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIWIRE NITINOL HYDROPHILIC WIRE GUIDE
Common Device Name
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13904986
Report Number1820334-2022-00461
Device Sequence Number4065849
Product Code OCY
UDI-Device Identifier00827002304758
UDI-Public(01)00827002304758(17)240404(10)71001590
Combination Product (Y/N)N
PMA/510(K) Number
K082536
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/04/2024
Device Model NumberG30475
Device Catalogue NumberHWAS-035150
Device Lot Number71001590
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-