TELEFLEX MEDICAL SDN. BHD. GIBECK HUMID-VENT FILTER SMALL STRAIGHT; FILTER, BACTERIAL, BREATHING-C Back to Search Results
Model Number IPN043782
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported "during use in patient, immediately upon replacing the filter, water condensation is observed, and activation of alarms in ventilator due to low supplied volume, this was corrected when the filter was changed for another".No patient injury or harm reported.Patient condition unknown at time of report.
 
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported "during use in patient, immediately upon replacing the filter, water condensation is observed, and activation of alarms in ventilator due to low supplied volume, this was corrected when the filter was changed for another".No patient injury or harm reported.Patient condition unknown at time of report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIBECK HUMID-VENT FILTER SMALL STRAIGHT
Common Device Name
FILTER, BACTERIAL, BREATHING-C
Manufacturer (Section D)
TELEFLEX MEDICAL SDN. BHD.
perak, west malaysia
Manufacturer (Section G)
TELEFLEX MEDICAL SDN. BHD.
lot no : pt2577 jalen perusahaan
4 kamunting industrial estate
perak, west malaysia 34600
MY   34600
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13905309
Report Number8040412-2022-00067
Device Sequence Number14469515
Product Code CAH
UDI-Device Identifier04026704348664
UDI-Public04026704348664
Combination Product (Y/N)N
Initial Reporter CountryMX
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2026
Device Model NumberIPN043782
Device Catalogue Number19502
Device Lot NumberKMZ21B0361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date05/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VENTILATOR
Patient SexUnknown
-
-